氮(dan)(dan)氣(qi)(qi)(qi)(qi)(qi)是(shi)zui常用(yong)的(de)(de)惰性(xing)氣(qi)(qi)(qi)(qi)(qi)體(ti),價格低廉(lian),易制(zhi)無毒(du),在實(shi)驗室中常用(yong)做色譜載氣(qi)(qi)(qi)(qi)(qi)、吹掃、保(bao)護等(deng)。實(shi)驗室的(de)(de)氮(dan)(dan)氣(qi)(qi)(qi)(qi)(qi)來源(yuan)主(zhu)要(yao)有(you)三種,一(yi)是(shi)鋼(gang)瓶(ping)氣(qi)(qi)(qi)(qi)(qi),二是(shi)管道氣(qi)(qi)(qi)(qi)(qi),三是(shi)氮(dan)(dan)氣(qi)(qi)(qi)(qi)(qi)發(fa)生(sheng)器(qi)(qi)。鋼(gang)瓶(ping)氣(qi)(qi)(qi)(qi)(qi)氣(qi)(qi)(qi)(qi)(qi)體(ti)質量(liang)高(gao)(gao),但鋼(gang)瓶(ping)屬(shu)于壓力容器(qi)(qi),運輸(shu)和保(bao)存需要(yao)一(yi)定的(de)(de)資質,偏(pian)遠地(di)區(qu)更換麻煩,費(fei)用(yong)高(gao)(gao);管道氣(qi)(qi)(qi)(qi)(qi)為(wei)大規模制(zhi)氮(dan)(dan),統(tong)一(yi)調(diao)度使用(yong),適(shi)合大型工(gong)(gong)廠或(huo)用(yong)氣(qi)(qi)(qi)(qi)(qi)單位集中的(de)(de)工(gong)(gong)業(ye)園區(qu),用(yong)氣(qi)(qi)(qi)(qi)(qi)量(liang)大建設費(fei)用(yong)高(gao)(gao);氮(dan)(dan)氣(qi)(qi)(qi)(qi)(qi)發(fa)生(sheng)器(qi)(qi)為(wei)現場制(zhi)氮(dan)(dan),多為(wei)小(xiao)型氣(qi)(qi)(qi)(qi)(qi)站或(huo)者實(shi)驗室儀器(qi)(qi)或(huo)小(xiao)型生(sheng)產線(xian)單*對(dui)一(yi)配套,使用(yong)靈活費(fei)用(yong)可(ke)控,對(dui)運輸(shu)和保(bao)存沒有(you)特殊要(yao)求,為(wei)越來越多的(de)(de)實(shi)驗室用(yong)戶選擇(ze)。變壓吸(xi)附(fu)(fu)空(kong)分(fen)制(zhi)氮(dan)(dan)是(shi)一(yi)種先進的(de)(de)氣(qi)(qi)(qi)(qi)(qi)體(ti)分(fen)離(li)技術,以進口碳分(fen)子(zi)篩為(wei)吸(xi)附(fu)(fu)劑,采用(yong)常溫下變壓吸(xi)附(fu)(fu)原理分(fen)離(li)空(kong)氣(qi)(qi)(qi)(qi)(qi)制(zhi)取高(gao)(gao)純度的(de)(de)氮(dan)(dan)氣(qi)(qi)(qi)(qi)(qi)。
由(you)于(yu)長期的使用,氮氣發生(sheng)器(qi)可(ke)(ke)能會出現故障,基(ji)本可(ke)(ke)按照以下的方法來解決:
一、開機不工作顯(xian)(xian)示(shi)(shi)板(ban)無顯(xian)(xian)示(shi)(shi)
1檢查電源是否有電;
2電路插(cha)件是否插(cha)牢;
3檢(jian)查(cha)保險(xian)是否燒(shao)掉(diao),保險(xian)位于儀器后面(mian)電線插(cha)座內(nei)(有(you)保險(xian)圖(tu)案(an)),用小“一”字改錐或(huo)尖的金屬工具(ju)向外撬即可取出(chu),保險(xian)為;
4打(da)開儀器側板(ban)檢查(cha)顯示板(ban)排線插(cha)件是否(fou)插(cha)牢,
5看(kan)儀器內電路(lu)板指示(shi)燈(deng)是否亮(liang),若不亮(liang),檢查(cha)電路(lu)板上保險是否燒掉,若燒掉須(xu)換保險3A,換后仍不顯示(shi),須(xu)換電源;
二、開機顯(xian)示“000”
檢查(cha)儀器后空氣(qi)輸入(ru)端(duan)是(shi)否有氣(qi)體(ti)輸入(ru)
三、開機10分(fen)鐘后(hou)顯示數字有但不升壓,無氣體(ti)輸出
檢查電(dian)(dian)(dian)磁閥(fa)(fa)插(cha)件是否(fou)脫落松動,用表(biao)對電(dian)(dian)(dian)磁閥(fa)(fa)這兩(liang)個插(cha)件進行測量(liang),若(ruo)沒有220V電(dian)(dian)(dian)壓需更換電(dian)(dian)(dian)源。
四(si)、運行過程中表頭顯示數字大壓力(li)達不到設定值(zhi)
是(shi)漏器造成的(de),對氣路進行(xing)全(quan)面(mian)檢漏特別是(shi)干燥室及(ji)電池。
五(wu)、檢查電池是否(fou)漏氣壞(huai)掉
1打開儀器側板,觀察電池板之間黑(hei)膠(jiao)墊看有無凸出變形和滲水現(xian)象,若(ruo)有即更換電池;
2使氮(dan)氣輸入5KG壓(ya)力后斷電,電池(chi)水路應(ying)是(shi)靜止(zhi)的若(ruo)有(you)水泡不斷冒出說明電池(chi)壞掉;
六、儀器運行中有響聲
是(shi)電(dian)磁(ci)閥響(xiang):用(yong)14的(de)(de)扳手對(dui)電(dian)磁(ci)閥上(shang)螺母適當調(diao)整松緊(jin)度,不要太緊(jin);若不行需(xu)拆開電(dian)磁(ci)閥對(dui)內(nei)部進行清洗(xi)(響(xiang)主要是(shi)因電(dian)磁(ci)閥內(nei)臟有雜質),若清洗(xi)完還不行,須更(geng)換(huan)新的(de)(de)。
七、開(kai)機即有氣(qi)體輸(shu)出
當剛開機壓(ya)力上升時須按一下前(qian)面(mian)的紅色延時開關,然后輸出(chu)(chu)壓(ya)力從輸出(chu)(chu)放掉等待10分鐘(zhong)后方可使用。